Retinal Photocoagulation Laser
Amplifying Medium | Nd:Gdvo |
Applications | Retinal Photocoagulation |
Pulse Duration | Nanosecond |
Ergonomics | Tabletop |
Treatment Beam
The tolerance between the actual output power and the pre-set value is ±15%.
Pulse duration: 0.01s to 3s ± 5%
Pulse interval: 0.05s to 1s ± 5%; or Single Pulse (SP)
Duty cycle1: When the output power is between 50mW and 500mW, the duty cycle is not restricted.Pulse Counter: 0 to 9,999
Continuously adjustable laser beam spot size: 50µm~500µm; allowable scale tolerance: ±15%; allowable minimum scale tolerance: ±15µm (50µm).
Beam divergence angle: 13.8°± 20% at the output of the adaptor
Laser power repeatability: RP ≤±10%Aiming Beam
Laser Type: SemiconductorCooling SystemSecondary Cooling System: Air-Cooled; Primary Cooling System: Intelligent TEC (thermal-electric-coupling) semi-conductor cooling.